A convenient method for testing and calibrating surface perfusion sensors has been developed. A phantom tissue model is used to simulate the nondirectional blood flow of tissue perfusion. A computational fluid dynamics (CFD) model was constructed in Fluent® to design the phantom tissue and validate the experimental results. The phantom perfusion system was used with a perfusion sensor based on clearance of thermal energy. A heat flux gage measures the heat flux response of tissue when a thermal event (convective cooling) is applied. The blood perfusion and contact resistance are estimated by a parameter estimation code. From the experimental and analytical results, it was concluded that the probe displayed good measurement repeatability and sensitivity. The experimental perfusion measurements in the tissue were in good agreement with those of the CFD models and demonstrated the value of the phantom tissue system.
Skip Nav Destination
Article navigation
October 2008
Research Papers
A Phantom Tissue System for the Calibration of Perfusion Measurements
Ashvinikumar V. Mudaliar,
Ashvinikumar V. Mudaliar
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Mechanical Engineering, Virginia Tech
, Blacksburg, VA 24061-0238
Search for other works by this author on:
Brent E. Ellis,
Brent E. Ellis
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Mechanical Engineering, Virginia Tech
, Blacksburg, VA 24061-0238
Search for other works by this author on:
Patricia L. Ricketts,
Patricia L. Ricketts
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Mechanical Engineering, Virginia Tech
, Blacksburg, VA 24061-0238
Search for other works by this author on:
Otto I. Lanz,
Otto I. Lanz
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Small Animal Clinical Sciences, Virginia Tech
, Blacksburg, VA 24061-0442
Search for other works by this author on:
Elaine P. Scott,
Elaine P. Scott
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Mechanical Engineering, Virginia Tech
; Department of Engineering, Seattle Pacific University
, Seattle, WA 98119-1957
Search for other works by this author on:
Thomas E. Diller
Thomas E. Diller
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Mechanical Engineering, Virginia Tech
, Blacksburg, VA 24061-0238
Search for other works by this author on:
Ashvinikumar V. Mudaliar
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Mechanical Engineering, Virginia Tech
, Blacksburg, VA 24061-0238
Brent E. Ellis
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Mechanical Engineering, Virginia Tech
, Blacksburg, VA 24061-0238
Patricia L. Ricketts
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Mechanical Engineering, Virginia Tech
, Blacksburg, VA 24061-0238
Otto I. Lanz
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Small Animal Clinical Sciences, Virginia Tech
, Blacksburg, VA 24061-0442
Elaine P. Scott
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Mechanical Engineering, Virginia Tech
; Department of Engineering, Seattle Pacific University
, Seattle, WA 98119-1957
Thomas E. Diller
Virginia Tech—Wake Forest University School of Biomedical Engineering and Sciences
; Department of Mechanical Engineering, Virginia Tech
, Blacksburg, VA 24061-0238J Biomech Eng. Oct 2008, 130(5): 051002 (10 pages)
Published Online: July 10, 2008
Article history
Received:
July 25, 2007
Revised:
January 19, 2008
Published:
July 10, 2008
Citation
Mudaliar, A. V., Ellis, B. E., Ricketts, P. L., Lanz, O. I., Scott, E. P., and Diller, T. E. (July 10, 2008). "A Phantom Tissue System for the Calibration of Perfusion Measurements." ASME. J Biomech Eng. October 2008; 130(5): 051002. https://doi.org/10.1115/1.2948417
Download citation file:
Get Email Alerts
Improvement in Active Cell Proliferation Area at Higher Permeability With Novel TPMS Lattice Structure
J Biomech Eng (November 2024)
Modeling Fatigue Failure of Cartilage and Fibrous Biological Tissues Using Constrained Reactive Mixture Theory
J Biomech Eng (December 2024)
A Numerical Study of Crack Penetration and Deflection at the Interface Between Peritubular and Intertubular Dentin
J Biomech Eng (December 2024)
Related Articles
Testing of a Noninvasive Probe for Measurement of Blood Perfusion
J. Med. Devices (March,2008)
Modeling and Estimating Simulated Burn Depth Using the Perfusion and Thermal Resistance Probe
J. Med. Devices (September,2013)
Temporal Resolution Phantom: A Device and Method for Measurement of a CT Scanner’s Temporal Resolution
J. Med. Devices (June,2008)
New Mathematical Model to Estimate Tissue Blood Perfusion, Thermal Contact Resistance and Core Temperature
J Biomech Eng (August,2012)
Related Chapters
Experimental results
Ultrasonic Methods for Measurement of Small Motion and Deformation of Biological Tissues for Assessment of Viscoelasticity
Conjugate Priors with Zero Occurrences: Analyst Beware! (PSAM-0435)
Proceedings of the Eighth International Conference on Probabilistic Safety Assessment & Management (PSAM)
Parameter Estimation of the Duffing Oscillator Using Poincaré Map and an Elitist Genetic Algorithm
Intelligent Engineering Systems through Artificial Neural Networks, Volume 20