A calibration phantom that can be used to measure the temporal resolution of a CT scanner was designed utilizing a deterministic design process. The system was first defined in terms of a set of functional requirements based on parameters of the imaging modality. It was necessary to generate multiple time-varying signals visible to the scanner, each with a pre-determined temporal frequency. Roll-off in the scanner’s ability to resolve the modulation of certain signals would be used to determine the scanner’s temporal resolution. Based on size limitations imposed by the scanning environment, the phantom utilizes multiple planetary gear assemblies, driven by a common shaft, to achieve a wide range of rotational velocities. Results obtained with an alpha prototype agreed with the theory. It was determined that further development of the phantom was necessary to increase the sensitivity of the measurement. The latest prototype phantom has been used to measure the temporal resolution of two different scanners and it was shown that temporal resolution of each is different from the gantry rotation time.
Skip Nav Destination
Article navigation
Design Of Medical Devices Conference Abstracts
Temporal Resolution Phantom: A Device and Method for Measurement of a CT Scanner’s Temporal Resolution
Alexander H. Slocum, Jr.,
Alexander H. Slocum, Jr.
MIT
, MGH, Cleveland Clinic
Search for other works by this author on:
Stephen E. Jones, MD,
Stephen E. Jones, MD
MIT
, MGH, Cleveland Clinic
Search for other works by this author on:
Rajiv Gupta, MD
Rajiv Gupta, MD
MIT
, MGH, Cleveland Clinic
Search for other works by this author on:
Alexander H. Slocum, Jr.
MIT
, MGH, Cleveland Clinic
Stephen E. Jones, MD
MIT
, MGH, Cleveland Clinic
Rajiv Gupta, MD
MIT
, MGH, Cleveland ClinicJ. Med. Devices. Jun 2008, 2(2): 027542 (1 pages)
Published Online: June 25, 2008
Article history
Published:
June 25, 2008
Citation
Slocum, A. H., Jr., Jones, S. E., and Gupta, R. (June 25, 2008). "Temporal Resolution Phantom: A Device and Method for Measurement of a CT Scanner’s Temporal Resolution." ASME. J. Med. Devices. June 2008; 2(2): 027542. https://doi.org/10.1115/1.2932429
Download citation file:
418
Views
Get Email Alerts
Cited By
Related Articles
Design of a Calibration Phantom for Measuring the Temporal Resolution of a Tomographic Imaging Device
J. Med. Devices (September,2007)
A Phantom Tissue System for the Calibration of Perfusion Measurements
J Biomech Eng (October,2008)
CT Visualization of Cryoablation in Pulmonary Veins
J. Med. Devices (June,2009)
Methods for Post Hoc Quantitative Computed Tomography Bone Density Calibration: Phantom-Only and Regression
J Biomech Eng (September,2018)
Related Proceedings Papers
Related Chapters
Dimensional Measurements and Calibration
Metrology and Instrumentation: Practical Applications for Engineering and Manufacturing
An Algorithm Improving Frequency Resolution Based on DDS Calibration
International Conference on Information Technology and Computer Science, 3rd (ITCS 2011)
Experimental results
Ultrasonic Methods for Measurement of Small Motion and Deformation of Biological Tissues for Assessment of Viscoelasticity